# Verifying and Maintaining Calibration
How to Verify Calibration
Here’s how to test whether Voltair II and your PPM Encoder are working together properly.
-
Calibrate Voltair II to the Encoder.
-
Take the Voltair II / Encoder out of your active airchain, but keep them wired together.
-
Feed Voltair II White Noise, at or slightly below nominal input level. In stereo installations, make sure the noise is going to both channels.
-
With this signal present:
-
The Energy meter should show a strong consistent level;
-
The Confidence meter should start painting full green bars on both Original and Processed displays;
-
The Decode Interval should be resetting every 4.8 seconds; and
-
The RTA should show a consistent level across its spectrum.
If this is not the case, try Calibrating again. If the problem persists, there may be trouble with your PPM Encoder or wiring.
Frequent Calibrations
If you need to calibrate Voltair II frequently, the Encoder might have problems. Check the Information tab on the Web interface to see if the last Calibration indicated an Encoder issue.
Switch to a backup Encoder, calibrate to the backup, and see if it’s more stable.
It is possible that your Encoder’s internal noise specs or clock may slowly change over time as hardware ages. Recalibrating your Voltair II and Encoder as part of periodic or annual maintenance regimen will do no harm.